Page 1189
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1173
Identifica-
tion
Description
.MR.MRS
Standard deviation for mean dosing rate for the whole
range in mL/min
.MR.MRT
Mean dosing rate for the whole range in mL/min.
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off)
.NMP
Number of measuring points in the measuring point list
.SLO
Electrode slope of the sensor used for the command (in
%)
.STY
Type of stop with which the command was stopped: 1
= normal;0 = manual or after error
.TITER
Titer value of the solution used for the command
.VOL
Dosed volume
5.6.6.3.4.2
DOS pH - General/Hardware
Tab: Method
▶ DOS pH ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the control device, the dosing device, the sen-
sor and the stirrer are defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead