4 properties - limit values – Metrohm tiamo 2.4 Manual User Manual
Page 1425
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.4
■■■■■■■■
1409
Cancel determination
The ongoing determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].
6.7.3.4
Properties - Limit values
Tab: Configuration
▶ Sensors ▶ [Edit] ▶ Properties... ▶ Limit values
NOTE
This tab only appears for sensors of the type pH electrode, ISE elec-
trode or Conductivity sensor.
Slope monitoring (for pH and ISE electrodes only)
on | off (Default value: off)
If this option is activated, then a check will be made to see whether the
Slope determined during calibration is within the defined limit values. If it
is outside them, a corresponding message appears and you can select
whether you will nevertheless accept the value or whether the previous
value is to be retained.
Lower limit
Lower limit value for the slope.
for pH electrode
Input range
–2.147E+9 - 2.147E+9 %
Default value
96.0 %
for ISE (+1)
Input range
–2.147E+9 - 2.147E+9 mV
Default value
55.0 mV
for ISE (–1)
Input range
–2.147E+9 - 2.147E+9 mV
Default value
–61.0 mV
for ISE (+2)
Input range
–2.147E+9 - 2.147E+9 mV
Default value
25.0 mV
for ISE (–2)
Input range
–2.147E+9 - 2.147E+9 mV
Default value
–31.0 mV