5 Method
tiamo 2.4
■■■■■■■■
1289
(case 1), the loop is run through. It is possible with certain stop conditions
that the loop is never run through.
NOTE
Nested and overlapping loops are not allowed.
Parameters
The parameters for the command LOOP are configured in the following
dialog window:
■
Command variables
The following command variables are generated in the method run by the
command LOOP and can be used in formulas under the designation
'Command name.Variable designation':
Identifi-
cation
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
not ever been started
.FIN
Command status; 1 = Command has been ended at
least once, 0, invalid (variable not available) = Com-
mand has never been ended
.LCO
Loop counter = current quantity of completed runs, both
of Repeat loops and of While loops
.LST
Start time of the loop command (Date/Time)
5.6.10.4.2
LOOP - Properties
Dialog window: Method
▶ LOOP ▶ Properties... ▶ LOOP - Loop #
Command name
Name of the command.
Entry
25 characters
Stop conditions
Definition of three possible stop conditions. The first stop condition to be
fulfilled will stop the loop.