4 det upol - titration parameters, Titration parameters – Metrohm tiamo 2.4 Manual User Manual
Page 566
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
550
■■■■■■■■
tiamo 2.4
5.6.3.2.5.4
DET Upol - Titration parameters
Tab: Method
▶ DET Upol ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the running of the titration are defined on this tab.
Titration rate
Titration rate
The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for the Measured value
acceptance and Dosing of increments of these sets are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.
Selection
slow | optimal | fast | User
Default value
optimal
slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted EPs.
optimal
Parameter set for all standard titrations; optimized for the most fre-
quent applications.
fast
For less critical rapid titrations.
User
Editing the individual titration parameters which affect the titration
rate.
Measured value acceptance
NOTE
Will be displayed only if the user option is selected in the selection list
Titration rate.
Measured value drift
Drift for the measured value acceptance during the titration.