7.7 814/815 USB Sample Processor
■■■■■■■■■■■■■■■■■■■■■■
1520
■■■■■■■■
tiamo 2.4
Swing offset
Swing angle offset for specific robotic arm model. This value must be set
according to the leaflet supplied with the swing head.
Input range
-270.0 - 270.0 °
Default value
0.0 °
Max. swing range
Maximum usable swing range for specific robotic arm model. Each robotic
arm model displays a different value on the basis of its construction. This
value must be set according to the leaflet supplied with the swing head
and can also be reduced as required.
Input range
0.0 - 330.0 °
Default value
60.0 °
Swing radius
The swing radius depends on the length of the robotic arm and, together
with the axial distance, is the most important parameter for exact move-
ment to a rack position. This value must be set according to the leaflet
supplied with the swing head.
Input range
30.0 - 300.0 mm
Default value
110.0 mm
Rotation offset
Offset from the middle of the tower to the middle of the robotic arm. If a
robotic arm must be mounted on the tower so that it is offset to one side,
then this value can be determined by the service technician during rack
adjustment.
Input range
-270.0 - 270.0 °
Default value
0.0 °
Swing direction
The swing direction of the robotic arm depends on the type of robotic
arm.
For a 2-tower model, the robotic arm at Tower 1 must swing to the right,
i.e. – and that at Tower 2 must be mounted + so that it swings to the left.
Selection
+ | –
Default value
+
+
Clockwise rotation.
–
Counterclockwise rotation.