7 met cond - additional evaluations, Additional evaluations – Metrohm tiamo 2.4 Manual User Manual
Page 703
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
687
Evaluation with volume window (mL)
An area (window) can be defined on the volume axis with this option.
The equivalence point is recognized only if it lies within this window
and in addition fulfills the parameters defined in the window.
Evaluation without window
Smoothing
Measure for the curve smoothing used for the data processing. The higher
the smoothing that is selected, the more the measurement curve will be
smoothed.
Input range
0 - 150
Default value
0
Evaluation with volume window (mL)
With this option up to 9 areas (windows) can be defined on the volume
axis. Only those equivalence points which lie within these windows and
additionally meet the parameters defined for each window will be recog-
nized. Only one equivalence point will be recognized per window. The
defined windows with their parameters are shown in the window table
and can be edited with the following buttons:
Table of the volume windows
The parameters for the defined volume window are displayed in the win-
dow table, which cannot be edited. The columns Lower limit mL,
Upper limit mL and Smoothing appear.
[New]
Opens the Volume window # dialog window, in which the parameters
for a new window can be entered.
[Properties]
Opens the Volume window # dialog window, in which the parameters
for the selected window can be edited (see Chapter 5.6.3.3.6.9, page
692).
[Delete]
Deletes the window selected in the table.
5.6.3.3.6.7
MET Cond - Additional evaluations
Tab: Method
▶ MET Cond ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Entry
25 characters