5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
832
■■■■■■■■
tiamo 2.4
Selection
Sensor name | Metal electrode
Default value
Metal electrode
I(pol)
Polarization current at the indicator electrode.
Selection
2 µA | 5 µA | 10 µA | 20 µA | 30 µA
Default value
10 µA
NOTE
When changing this parameter, make sure that the polarization current
of the indicator electrode, the endpoint and the control range are
linked closely with one another (see Chapter 5.6.3.6.4, page 834).
Electrode check
on | off (Default value: on)
If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status
to a measurement. A check is also made while doing so to ensure that the
electrode is properly connected and that no short-circuit is present.
Temperature measurement
Type of temperature measurement.
851, 852
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will be used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Cell
Generator type
Selection of the generator type.