Metrohm tiamo 2.4 Manual User Manual
Page 767
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
751
Max. rate
Rate at which dosing is carried out outside of the control range. The
maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.
Titrando, 855
Input range
0.01 - 166.00 mL/min
Default value
1.00 (slow), 10.00 (optimal), 166.00
(fast) mL/min
Selection
maximum
Titrino
Input range
0.01 - 150.00 mL/min
Default value
1.00 (slow), 10.00 (optimal), 150.00
(fast) mL/min
Selection
maximum
Min. rate
Rate at which dosing is carried out at the very beginning of the titration
and in the control range at the end of the titration. The smaller the
selected minimum rate, the slower the titration.
Input range
0.01 - 9999.00 µL/min
Default value
5.00 (slow), 25.00 (optimal), 50.00
(fast) µL/min
Stop criterion
Stop criterion
The titration us stopped after the last dosing once the endpoint has been
reached and the stop criterion has been fulfilled. The titration can be stop-
ped when a particular drift has been achieved or after a preset time.
NOTE
The Stop conditions will always stop the titration even if the stop crite-
rion has not been fulfilled or has been turned off.
Selection
drift | time | off
Default value
drift
drift
The switch-off point will always have the same curve slope dV/dt.