Metrohm tiamo 2.4 Manual User Manual
Page 713
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
697
Identification
Description
.CBY
Command status; 1 = Conditioning activated, 0 =
Conditioning deactivated
.COK
Command status; 1 = Condition requirement ful-
filled, 0 = Condition requirement not fulfilled
.CONC
Concentration of the solution used for the command
.CYL
Cylinder volume of the exchange or dosing unit used
for the command
.DBL
Total duration for the processing of the command in
s
.DRI
Current and/or last drift for drift correction in µL/min
.DSC
Time for processing all start conditions in s
.DTI
Time in s for drift correction (time from the start of
the titration to the end of the command)
.EME
End measured value (measured value after processing
of the command) in the unit of the measured value
.ENP
Electrode zero point of the sensor used for the com-
mand (dimensionless)
.EP{x}.MEA
Measured value for the endpoint × (1 - 9) in the unit
of the measured value
.EP{x}.MEP
Number of endpoints in the window × (1 - 9); 1 = 1
endpoint, 2 = 2 or more endpoints, 3 = EP cor-
rected with Autodrift, 4 = EP corrected with
manual drift
.EP{x}.TEM
Temperature for the endpoint × (1 - 9) in °C
.EP{x}.TIM
Time in s until the endpoint × (1 - 9) is reached
.EP{x}.VOL
Volume for the endpoint × (1 - 9) in mL
.ETE
End temperature (temperature after the command
has been processed) in °C
.EVT
End volume (total dosed volume at the end of the
command) in mL
.FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) =
Command has never ended
.FP{x}.MEA
Measured value for the fixed endpoint × (1 - 9) in mV