5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
754
■■■■■■■■
tiamo 2.4
auto
The titration direction is determined automatically from the start mea-
sured value and the set endpoint.
Extraction time
The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). For example, it makes sense to
enter an extraction time for the titration of samples that are not easily
soluble.
Titrando, 719, 720, 736, 751, 758, 785, 794, 798, 799
Input range
0 - 999999 s
Temperature
Manually entered titration temperature. If a temperature sensor is con-
nected and the Temperature measurement is set on the General/
Hardware tab under Sensor to automatic or continuous, then the
temperature will be measured continuously. This value is used for temper-
ature correction in pH measurements.
Titrando, 855
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
Titrino
Input range
-170.0 - 500.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Titrando, 855
Input range
0.1 - 999999 s
Default value
2.0 s
Titrino
Input range
0.08 - 16200 s
Default value
2.00 s
5.6.3.4.5.6
SET Ipol - Stop conditions
Tab: Method
▶ SET Ipol ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Entry
25 characters