2 cal spec - general/hardware – Metrohm tiamo 2.4 Manual User Manual
Page 1127
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1111
Identification
Description
.C3
Coefficient c3 of the calibration curve
.CAL
Status of the calibration: 1 = calibrated normally,
0 = canceled, because reference wavelength was
not found.
.DBL
Total duration for the processing of the command in
s
.STY
Type of stop with which the command was stopped:
1 = stopped normally, 0 = stopped manually
with [Quit] or [Stop] .
.CP{#}.MEA
Measured wavelength of the calibration point in nm.
If no measured peak can be assigned, then not
found will be entered.
.CP{#}.REF
Reference wavelength of the calibration point in nm.
.FIN
Command status: 1 = Command has been ended
at least once, 0 = Command was started but
not ended, invalid = Command was not started
(variable not available).
.TOL
Tolerance within which a measured peak can be
assigned to a calibration wavelength.
5.6.5.5.2
CAL Spec - General/Hardware
Tab: Method
▶ CAL Spec ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the instrument are defined on this tab.
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.
In two-channel measurements, i.e. in a master/slave setup, only master
spectrometers can be selected.
In one-channel measurements, i.e. no master/slave setup, only instru-
ments for which the entry not defined is selected under Device assign-
ment in the configuration can be selected.