6 Configuration
tiamo 2.4
■■■■■■■■
1415
ELT interval
Time interval to next electrode test. If a value is entered here, then the
date in the Next ELT test field will be adjusted automatically.
Only editable for Monitoring ELT validity = on, otherwise inactive.
Input range
1 - 999 days
Default value
999 days
Next ELT test
Date on which the next electrode test is to be carried out. The date can be
selected by clicking on
in the Select date dialog window. After the
date has been entered, the ELT interval field will be adjusted automati-
cally.
Only editable for Monitoring ELT validity = on, otherwise inactive.
Default value
Date ELT+ 999 days
Message
Message by e-mail
on | off (Default value: off)
If this check box is activated, the message is sent to the address defined
under [E-Mail...]. The message is sent in text format.
Only editable for Monitoring ELT validity = on, otherwise inactive.
[E-mail...]
Opens the Send e-mail window (see Chapter 2.6.1, page 93).
Only active if Message by e-mail = on, otherwise inactive.
Acoustic signal
on | off (Default value: off)
If this check box is activated, an acoustic signal will be emitted in addition
to the message.
Only editable for Monitoring ELT validity = on, otherwise inactive.
Action
If it is found during monitoring that the validity period has expired, then
one of the following actions will be triggered automatically during the
start test:
Only editable for Monitoring ELT validity = on, otherwise inactive.