5 Method
tiamo 2.4
■■■■■■■■
953
Command variables
The following command variables are generated in the method run by the
command MEAS U and can be used in formulas under the designation
'Command name.Variable designation':
Designation
Description
.BLV
Blank value of the sensor used for the command (only
for ISE sensors)
.BP{x}.MEA
Measured value for the break point x (1…9) in the unit
of the measured value
.BP{x}.TEM
Temperature for the break point x (1…9) in °C
.BP{x}.TIM
Time for the break point x (1…9) in s
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.DBL
Total duration for the processing of the command in s
.EME
End measured value (measured value after processing
of the command) in the unit of the measured value
.ENP
Electrode slope of the sensor used for the command (in
mV for ISE sensors)
.ETE
End temperature (temperature following processing of
the command) in °C
.FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) = Com-
mand has never ended
.FP{x}.MEA
Measured value for the fixed endpoint x (1…9) in the
unit of the measured value
.FP{x}.TEM
Temperature for the fixed endpoint x (1…9) in °C
.FP{x}.TIM
Time in s until the fixed endpoint x (1…9) is reached
.IME
Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.CAx
Calculated value x (1…3) for the last measuring point
on the measuring point list