3 stdadd dos - standard addition – Metrohm tiamo 2.4 Manual User Manual
Page 1099
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1083
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
under Temperature on the Measuring parameters tab will be
used.
off
The temperature will not be measured. The temperature manually
entered under Temperature on the Measuring parameters tab will
be used.
Stirrer
Stirrer
Selection of the stirrer.
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Input range
-15 - 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically
when the command has finished.
5.6.4.18.3
STDADD dos - Standard addition
Tab: Method
▶ STDADD dos ▶ Properties... ▶ Standard addition
Command name
Name of the command.
Entry
25 characters
The parameters for the standard addition can be defined on this tab
Number of additions
Number of times the standard solution is added.
Input range
1 - 19
Default value
3