Metrohm tiamo 2.4 Manual User Manual
Page 833
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
817
has been reached. The larger the control range, the slower the titration.
Outside the control range, dosing is carried out continuously at Max.
rate.
Titrando
Input range
0.01 - 200.0 µA
Default value
40.0 (slow), 10.0 (optimal), 5.0 (fast) µA
Selection
off
off
The entire measured value range is defined as control range. The mini-
mum titration rate is applied throughout the entire titration.
Titrino
Input range
0.1 - 200.0 µA
Default value
40.0 (slow), 10.0 (optimal), 5.0 (fast) µA
Selection
off
off
The entire measured value range is defined as control range. The mini-
mum titration rate is applied throughout the entire titration.
Max. rate
Rate at which dosing is carried out outside of the Control range. The
maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.
Titrando
Input range
0.01 - 166.00 mL/min
Default value
1.00 (slow), maximum (optimal), maximum
(fast) mL/min
Selection
maximum
Titrino
Input range
0.01 - 150.00 mL/min
Default value
1.00 (slow), maximum (optimal), maximum
(fast) mL/min
Selection
maximum
Min. volume increment
This parameter determines the addition rate with which dosing is carried
out at the very beginning and at the end of the titration. The smaller the
increment that is selected, the slower the titration and the higher the
accuracy.
Input range
0.1 - 9.9 µL
Default value
minimum (slow), minimum (optimal), 5.0
(fast) µL