6.6 Subwindow Titrants/Solutions
■■■■■■■■■■■■■■■■■■■■■■
1386
■■■■■■■■
tiamo 2.4
Cylinder serial number
Serial number of the cylinder. It is printed on new cylinders. With intelli-
gent exchange units it is read out automatically. It can be modified at any
time, e.g. if the cylinder has to be replaced.
IEU
Entry
8 characters
Default value
'empty'
EU
Entry
10 characters
Default value
'empty'
Parameters for preparation
Configuration of the parameters which are to be used in the PREP com-
mand.
Volume
Volume to be dosed in during preparation.
Input range
0.00000 - 99,999.99999 mL
Selection
Cylinder volume
Default value
Cylinder volume
Cylinder volume
The whole contents of the cylinder will be dosed.
Cycles
Number of rinsing cycles during preparation.
Input range
1 - 9
Default value
2
Dosing rate
Rate at which the solution is to be dosed. The maximum dosing rate
depends on the cylinder volume of the exchange unit used (see below).
When the function is carried out, the rate will automatically be reduced to
the highest possible value.
Input range
0.01 - 150.00 mL/min
Selection
maximum
Default value
maximum
Filling rate
Rate at which the cylinder is to be filled. The maximum filling rate depends
on the cylinder volume of the exchange unit used (see below). When the