6 det ipol - potentiometric evaluation, Potentiometric evaluation – Metrohm tiamo 2.4 Manual User Manual
Page 544
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
528
■■■■■■■■
tiamo 2.4
Titrando, 855
Input range
0 - 999999 s
Selection
off
Default value
off
Filling rate
Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.
Titrando, 855
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Input range
0.01 - 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.2.4.6
DET Ipol - Potentiometric evaluation
Tab: Method
▶ DET Ipol ▶ Properties... ▶ Potentiometric evaluation
Command name
Name of the command.
Entry
25 characters
Parameters for the potentiometric evaluation of the titration curve with
automatic equivalence point recognition are defined on this tab. One of
the following three evaluation methods can be chosen here:
Potentiometric evaluation
Selection
Evaluation without window | Evaluation with
measured value window (Upol) | Evaluation
with volume window (mL)
Default value
Evaluation without window
Evaluation without window
With this option set, the two parameters EP criterion and EP recog-
nition will be applied across the entire range of the titration curve.