5 Method
tiamo 2.4
■■■■■■■■
827
[New]
Opens the External measured value # dialog window, in which the
parameters for the external measured value can be entered.
[Properties]
Opens the External measured value # dialog window, in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 934).
[Delete]
Deletes the external measured value selected in the table.
5.6.3.5.3.10
KFT Upol - Fixed endpoint evaluation
Dialog window: Method
▶ KFT Upol ▶ Properties... ▶ Additional evalua-
tions
▶ Fixed endpoint evaluation ▶ [New/[Properties] ▶ Fixed endpoint evalua-
tion #
Quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value
Input range
-200.0 - 200.0 µA
Time
Input range
0.0 - 999999.9 s
Volume
Input range
0.00000 - 9999.99 mL
5.6.3.6
KFC
5.6.3.6.1
KFC - Overview
Dialog window: Method
▶ KFC ▶ Properties.. ▶ KFC - 'Command name'
Command for coulometric Karl Fischer Titrations with voltametric mea-
surement.