5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1214
■■■■■■■■
tiamo 2.4
Dosing device
Dosing device
Selection of the dosing device to be prepared. All the dosing device con-
nections which are possible with the selected device type are displayed.
778, 789, 814, 815, 855, 864, 874
Selection
1 | 2 | 3
Default value
1
730, 774
Selection
1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12
Default value
1
Solution
Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent dosing units are used, then a
check will be made in the method run to see whether the correct solution
has been attached to the selected dosing device. With non-intelligent dos-
ing units, only the cylinder volume is checked. At the start of the com-
mand, a check is made of the working life, the validity of the titer and the
GLP test interval for the selected solution.
Entry
24 characters
Selection
'Solution name' | not defined
Default value
not defined
not defined
No tests will be carried out.
NOTE
For Dosing device and Solution exactly the same settings have to be
selected as for the dosing device reserved earlier by another command,
otherwise the dosing device is not being released.
NOTE
If the tandem dosing is used in a command (e.g. ADD), then both dos-
ing devices used have to be released each with a separate command
RLS DOS.