5 Method
tiamo 2.4
■■■■■■■■
1191
Titrando, 855
Selection
Solution name | not defined
Default value
not defined
not defined
No tests will be carried out.
Filling rate
Speed with which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the
exchange unit or dosing device used. If the entered filling rate is too high
for the selected dosing device, it will automatically be reduced during dos-
ing to the largest possible value.
Titrando, 855
Input range
0.01 - 166.00 mL/min
Default value
maximum mL/min
Sensor
Measuring input
Selection of the measuring input to which the sensor is connected.
Titrando
Selection
1 | 2
Default value
1
855
Selection
1
Default value
1
Titrino
Selection
1 | 2 | diff.
Default value
1
Sensor
Selection of a sensor of type pH electrode, metal electrode, pH elec-
trode or ISE electrode from the list of sensors defined in the sensor
table. The calibration data for the sensor will be adopted for pH and ISE
electrodes.
Selection
Sensor name | pH electrode | Metal electrode |
ISE electrode
Default value
Metal electrode
Temperature measurement
Type of temperature measurement.