2 login / password protection, Login/password protection – Metrohm tiamo 2.4 Manual User Manual
Page 1329
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.4
■■■■■■■■
1313
Security settings can be exported and imported. This means that these set-
tings can be exchanged between different client/server systems.
The security settings can be output as a PDF file with the File
(PDF)...
▶ Security settings menu item.
Settings according to 21 CFR Part 11
Settings according to 21 CFR Part 11
on | off (Default value: off)
If this check box is activated, then the default settings pursuant to 21 CFR
Part 11 will be activated on all tabs. At the same time, those check boxes
which are required to be activated pursuant to 21 CFR Part 11 will be dis-
played against a gray background, i.e. they can no longer be edited.
Tabs
The security settings are configured on the following tabs:
■
Login / password protection
Settings for login and password protection
■
Audit Trail / Changes
Activation/deactivation of the Audit Trail and comments when meth-
ods, determinations or sample data are modified
■
Signatures
Options for electronic signatures
■
Default reasons
Definition of reasons for signing and modifying methods, determina-
tions and sample data.
6.2.2.2
Login / password protection
Tab: Configuration
▶ Tools ▶ Security settings ▶ Login/Password protection
Definition is made on the Login/Password protection tab as to
whether the user must log in with name only or with name and password
and how the password must be monitored and constructed.
Login policies
Enforce login with user name
If this check box is activated, then each time the program starts the login
window will appear, in which the user must enter his or her user name. If
this check box is deactivated, no login is necessary and the user logged
in under Windows will be adopted as the user name. If this is the case,
then all subsequent parameters are deactivated.
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)