5 Method
tiamo 2.4
■■■■■■■■
807
Parameters
The parameters for the KFT Upol command are set on the following
eight tabs:
■
General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.
■
Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.
■
Control parameters
Control parameters for the endpoint (measured value for the endpoint,
titration rate, user-defined parameters, stop criterion).
■
Titration parameters
Parameters for the run of the titration.
■
Stop conditions
Definition of conditions which cause the titration to stop.
■
Conditioning
Enable/disable conditioning. Definition of start drift, drift correction
and stop conditions.
■
Additional evaluations
Definition of additional methods for the evaluation of the titration
curves.
■
Additional measured values
Definition of additional measured values of other measuring com-
mands, which are saved as additional columns in the measuring point
list.
Command variables
The following command variables are generated by the KFT Upol com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':
Identification
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.CBY
Command status; 1 = Conditioning activated, 0 =
Conditioning deactivated
.COK
Command status; 1 = Condition requirement ful-
filled, 0 = Condition requirement not fulfilled
.CONC
Concentration of the solution used for the command