6.7 Sensors subwindow
■■■■■■■■■■■■■■■■■■■■■■
1406
■■■■■■■■
tiamo 2.4
Input range
–2,000.0 - 2,000.0 mV
Default value
0.0 mV
c (blank)
c (blank) is the third characteristic of the calibration function U = f(log
c). It reflects to a certain extent the curvature of the calibration function at
the lowest concentrations. This curvature is caused by the influence of so-
called interfering ions. If the concentration calibration is carried out with
less than 3 standards, then c (blank) will be set equal to zero. The c
(blank) parameter is available for calculations as the variable 'Command
name.BLV'.
Value
Input range
0.00 - 999,999,999
Default value
0.00
Unit
Entry
10 characters
Default value
'empty'
Selection
mol/L | mmol/L | ppm | % | g/L | mg/L | µg/L |
mg/mL | µg/mL
Specific calibration data for conductivity sensors
Cell constant
Cell constant of the conductivity sensor. This value can be determined
automatically with a calibration or entered manually. The cell constant is
available for calculations as the variable 'Command name.BLV'.
Input range
0.001 - 500 /cm
Default value
1 /cm
Specific calibration data for the Thermoprobe
a ... e
Display of the calibration coefficients a … e.
Shared calibration data
Calibration temperature
Temperature during the calibration. In addition, the display shows
whether the temperature was measured with a Pt1000 or an NTC temper-
ature sensor or entered manually (is not being displayed for the Thermop-
robe).
Input range
–20.0 - 150.0 °C
Default value
25.0 °C