3 meas cond - measuring parameters, Measuring parameters – Metrohm tiamo 2.4 Manual User Manual
Page 1038
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1022
■■■■■■■■
tiamo 2.4
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Input range
-15 - 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is activated, the stirrer will be switched off automatically
when the command has finished.
5.6.4.10.3
MEAS Cond - Measuring parameters
Tab: Method
▶ MEAS Cond ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement
Measurement frequency
712
Selection
auto | 300 Hz | 2.4 kHz
Default value
auto
auto
The more suitable of the two frequencies is selected.
Type of measurement
The two measuring modes Measurement with drift control or Mea-
surement without drift control can be selected.
Selection
Measurement with drift control | Measure-
ment without drift control
Default value
Measurement with drift control
Measurement with drift control
The measurement is carried out drift-controlled if this option is
selected. The measurement is canceled as soon as the defined Signal
drift or the Stop measured value is reached or a Waiting time has
elapsed. The corresponding parameters are not visible if this option is
disabled. This option is not available for the 712 Conductometer.