3 rack properties, 1 editing rack properties, 2 properties - rack parameters – Metrohm tiamo 2.4 Manual User Manual
Page 1479: Rack properties 3
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.4
■■■■■■■■
1463
6.12.3
Rack properties
6.12.3.1
Editing rack properties
Dialog window: Configuration
▶ Rack data ▶ [Edit] ▶ Properties… ▶ Rack data
Rack name
Shows the name of the rack attached. If no rack is in position then "-----"
will be shown.
Rack code
Shows the number of positions on the rack in position. The rack code cor-
responds to the arrangement of magnets on the base of the rack and is
read in by the Sample Processor in order to recognize the rack. If no rack
is in position then the display will be empty.
Number of positions
Shows the number of positions on the rack. If no rack is in position then
the display will be empty.
The parameters for the rack in position are defined in the following tabs:
■
Rack parameters
Defines the parameters that are valid for all rack positions.
■
Lift positions
Defines the work, rinse, shift and special positions for Tower 1 and
Tower 2 (if present).
■
Special beakers
Special beaker Settings for all special beakers on the rack in position.
6.12.3.2
Properties - Rack parameters
Tab: Configuration
▶ Rack data ▶ [Edit] ▶ Properties... ▶ Rack data ▶ Rack
parameters
Beaker radius samples
Radius of the sample beakers on the rack. If the lift is moved to the work
position then this value will be compared with the parameter Min.
beaker radius (see Chapter 7.3.3, page 1484) that can be specifically
defined for each tower. If Beaker radius samples < min. beaker
radius then a corresponding error message will appear. With off the
beaker radius will not be checked.
Input range
1.0 - 100.0 mm
Selection
off
Beaker sensor
When a sample position is moved to with the command MOVE then the
beaker sensor (Tower, Robotic arm) checks whether the special beaker