2 kft ipol - general/hardware, General/hardware – Metrohm tiamo 2.4 Manual User Manual
Page 804
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
788
■■■■■■■■
tiamo 2.4
Identification
Description
.MA.VOL
Volume at maximum measured value in mL
.MI.MEA
Minimum measured value in the unit of the measured
value
.MI.TEM
Temperature at minimum measured value in °C
.MI.TIM
Time for the minimum measured value in s
.MI.VOL
Volume at minimum measured value in mL
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off)
.NMP
Number of measuring points in measuring point list
.SME
Start measured value (measured value after process-
ing the start conditions) in the unit of the measured
value
.STE
Start temperature (temperature after processing the
start conditions) in °C
.STY
Type of stop with which the command was stopped:
1 = normal; 0 = manual or after error
.SVA
Start volume absolute (volume that was added
according to the start condition "start volume") in mL
.SVT
Total start volume (volume that was added according
to all three start conditions) in mL
.TITER
Titer value of the solution used for the command
5.6.3.5.2.2
KFT Ipol - General/Hardware
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the instrument, the dosing device, the sensor
and the stirrer are defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.