5 meas t/flow - additional measured values, Additional measured values – Metrohm tiamo 2.4 Manual User Manual
Page 1024
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1008
■■■■■■■■
tiamo 2.4
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated time for the minimum mea-
sured value will be interpolated from the measuring point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
0.1 - 20.0 °C/s
Default value
1.0 °C/s
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is enabled, then the associated time for the maximum mea-
sured value will be interpolated from the measuring point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
0.1 - 20.0 °C/s
Default value
1.0 °C/s
5.6.4.8.5
MEAS T/Flow - Additional measured values
Tab: Method
▶ MEAS T/Flow ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Entry
25 characters
A maximum of 6 additional measured values can be defined on this tab
which can then be saved together with the measured values present in
the default settings in additional measured value columns Calculated #
or External # in the measuring point list .
Additional calculated measured values
Additional calculated measured values
on | off (Default value: off)
If this option is enabled, then a maximum of 3 new measured values can
be calculated from existing measured values and saved, which can also be