Metrohm tiamo 2.4 Manual User Manual
Page 538
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
522
■■■■■■■■
tiamo 2.4
Selection
off
Default value
off
Dosing rate
Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the
exchange unit or dosing unit used.
Input range
0.01 - 166.00 mL/min
Default value
5.00 mL/min
Selection
maximum
Start slope
NOTE
Only displayed for Titrando and 855.
Start slope
When the start slope is reached the dosing of the start volume is stopped
and the titration is started. If the start slope is achieved by the dosing of a
start volume then the titration starts directly.
Input range
0 - 999 mV/mL
Selection
off
Default value
off
Dosing rate
Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the
exchange unit or dosing unit used.
Input range
0.01 - 166.00 mL/min
Default value
5.00 mL/min
Selection
maximum
Pause
Pause
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.
Input range
0 - 999999 s
Default value
0 s