5 met ipol - stop conditions, Stop conditions – Metrohm tiamo 2.4 Manual User Manual
Page 650
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
634
■■■■■■■■
tiamo 2.4
Selection
maximum
Default value
maximum
Titrino
Input range
0.01 - 150.00 mL/min
Selection
maximum
Default value
maximum
Temperature
Temperature
Manually entered titration temperature. If a temperature sensor is con-
nected and the Temperature measurement is set on the General/
Hardware tab under Sensor to automatic or continuous, then the
temperature will be measured continuously. This value is used for temper-
ature correction in pH measurements.
Titrando, 855
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
Titrino
Input range
-170.0 - 500.0 °C
Default value
25.0 °C
5.6.3.3.4.5
MET Ipol - Stop conditions
Tab: Method
▶ MET Ipol ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Entry
25 characters
Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.
Stop volume
Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.
Titrando, 855
Input range
0.00000 - 9999.99 mL
Default value
100,000 mL
Selection
off