Metrohm tiamo 2.4 Manual User Manual
Page 1213
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1197
Lower hysteresis
Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value
lower limit ok is triggered.
Titrando, 855
Input range
0.0 - 2000.0 mV
Default value
2.0 mV
Upper limit
Upper limit of the measured value. If the measured value exceeds this
limit, then the event Measured value upper limit exceeded is trig-
gered.
Titrando, 855
Input range
-2000.0 - 2000.0 mV
Default value
2000.0 mV
Titrino
Input range
-2000 - 2000 mV
Default value
2000 mV
Upper hysteresis
The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.
Titrando, 855
Input range
0.0 - 2000.0 mV
Default value
2.0 mV
Action
Selection of the event to be triggered if the upper or lower limit is
exceeded:
Selection
Stop determination | Stop command (only
Titrando) | Wait for [Continue] | Wait for limit
ok | none
Default value
none
Stop determination
The running DOS command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running DOS command is quit, then the next command is exe-
cuted.