5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
752
■■■■■■■■
tiamo 2.4
time
The endpoint must be exceeded for a certain time: the Delay time. The
same Delay time with different very small volume increments (depend-
ing on the volume of the exchange or dosing unit) means different
switch-off points.
off
The titration will not be stopped. In older instructions the Delay time
was usually defined as the Stop criterion.
Stop drift
Titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift.
Input range
1 - 999 µL/min
Default value
20 µL/min
Delay time
If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time. The selection inf. stands for infin-
ite.
Titrando, 855
Input range
0 - 999 s
Default value
10 s
Titrino
Input range
0 - 999 s
Default value
10 s
Selection
inf.
Stop time
The titration is stopped if the stop time defined here has elapsed since the
start of the titration. This parameter is only shown for Stop criterion =
time and Delay time = inf.
Titrino
Input range
0 - 999999 s
Selection
off
Default value
off