5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1182
■■■■■■■■
tiamo 2.4
Titrando, 855
Input range
-20.000 - 20.000
Default value
20.000
Titrino
Input range
-20.00 - 20.00
Default value
20.00
Upper hysteresis pH
The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.
Titrando, 855
Input range
0 - 20.000
Default value
20
Action
Selection of the event to be triggered if the upper or lower limit is
exceeded:
Selection
Stop determination | Stop command (only
Titrando) | Wait for [Continue] | Wait for limit
ok | none
Default value
none
Stop determination
The running DOS command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running DOS command is quit, then the next command is exe-
cuted.
Wait for [Continue]
Reagent dosing in the current DOS command is interrupted and a
message appears. As soon as the monitored measured value is again
within the limits (including hysteresis) reagent dosing can be resumed
by pressing the [Continue] button in this message box.
Wait for limit ok
Reagent dosing in the current DOS command is interrupted. As soon
as the monitored measured value is again within the limits (including
hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.