6 Configuration
tiamo 2.4
■■■■■■■■
1347
Min. offset potential
Minimum permitted offset voltage (voltage at pH = 7.0). This value applies
for all evaluations.
Input range
-999 - 999 mV
Default value
-15 mV
Max. offset potential
Maximum permitted offset voltage (voltage at pH = 7.0). This value
applies for all evaluations.
Input range
-999 - 999 mV
Default value
15 mV
6.3.3.6.5
Usable electrode
Tab: Configuration
▶ Tools ▶ Templates ▶ Electrode types... ▶ [Properties] ▶ Usa-
ble electrode
Max. streaming potential
Maximum permitted flow potential (difference of the measured voltages
in stirred and non-stirred solution).
Input range
0.0 - 999.9 mV
Default value
4.0 mV
Max. drift
Maximum permitted drift in stirred solution.
Input range
0.1 - 9.9 mV/min
Default value
3.0 mV/min
Min. slope
Minimum permitted slope of the pH electrode.
Input range
-999.9 - 999.9 %
Default value
95.0 %
Max. slope
Maximum permitted slope of the pH electrode.
Input range
-999.9 - 999.9 %
Default value
103.0 %
Max. response time
Maximum permitted response time of the pH electrode. The voltage mea-
sured after three minutes in stirred solution is used as a comparison value
for determining the response time. The response time is the time after