5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1164
■■■■■■■■
tiamo 2.4
Titrando, 846, 814, 815, 855, 864, 874
Input range
0.00000 - 99,999.9 mL
Default value
10.0000 mL
Titrino
Input range
0.0001 - 9,999.0 mL
Default value
10.0000 mL
730, 778, 789
Input range
0.001 - 999.999 mL
Default value
10.0000 mL
774
Input range
0.01 - 999.999 mL
Default value
10.0000 mL
Dosing rate
Rate (volume per time unit) at which the volume is dosed. The maximum
dosing rate depends on the cylinder volume of the exchange unit or dos-
ing device used. If the entered dosing rate is too high for the selected dos-
ing device, it will automatically be reduced during dosing to the largest
possible value.
Titrando, 846, 814, 815, 855, 864, 874
Input range
0.01 - 166.00 mL/min
Default value
maximum mL/min
730, 774, 778, 789
Input range
0.01 - 160.00 mL/min
Default value
maximum mL/min
Titrino
Input range
0.01 - 150.00 mL/min
Default value
maximum mL/min
NOTE
The dosing rate should be reduced for viscous liquids.
Filling rate
The buret will be filled at this rate after dosing. The maximum filling rate
depends on the cylinder volume of the exchange unit or dosing device
used. If the entered filling rate is too high for the selected dosing device, it
will automatically be reduced during filling to the largest possible value.