5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1078
■■■■■■■■
tiamo 2.4
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Input range
0.1 - 999999.0 s
Default value
2.0 s
Temperature
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.
Input range
-20.0 - 150°C
Default value
25.0
Switch off stirrer during measurement
Switch off stirrer during measurement
on | off (Default value: off)
If this option is activated, then the stirrer is switched off during the mea-
surement. In this case the following two parameters can be edited:
Stir before measurement
Stirring time between end of dosing and measurement.
Input range
0 - 999999 s
Default value
0 s
Pause before measurement
Waiting time before the measurement during which no stirring takes
place.
Input range
0 - 999999 s
Default value
0 s
5.6.4.18
STDADD dos
5.6.4.18.1
STDADD dos - Overview
Dialog window: Method
▶ STDADD dos ▶ Properties... ▶ STDADD dos - 'Com-
mand name'
Command for Standard addition with addition of the standard addition
solution from a dosing device.
Devices