4 meas cond - evaluations, Evaluations – Metrohm tiamo 2.4 Manual User Manual
Page 1042
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1026
■■■■■■■■
tiamo 2.4
Sample solution
Selection of a function for the temperature coefficient from the table
Sample solutions (TC conductivity).
Temperature coefficient
The temperature compensation is calculated on the basis of the value
entered (constant).
Input range
0.00 - 9.99 %/°C
Default value
2.00 %/°C
Sample solution
The temperature compensation is carried out on the basis of temperature-
dependent coefficients and should be used for determinations where the
sample temperature is not constant. There are two possibilities:
■
It is possible to use a sample solution whose temperature coefficient
has been determined by a previous measurement (MEAS TC Cond).
■
For ground water, spring water and surface water, the temperature
coefficients according to standard ISO 7888:1985 (German version:
DIN EN 27888:1993) are stored in the system.
Selection
DIN | Table entries
Default value
DIN
5.6.4.10.4
MEAS Cond - Evaluations
Tab: Method
▶ MEAS Cond ▶ Properties... ▶ Evaluations
Command name
Name of the command.
Entry
25 characters
The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.
Fixed endpoint evaluation
Fixed endpoint evaluation
on | off (Default value: off)
If this option is activated, then the associated values for the other mea-
sured quantities will be interpolated with a fix measured quantity (Mea-
sured value or Time) for the fixed endpoint from the measuring point
list.
Moves the selected line up (changes the sequence).