The conductivity
measured at a particular temperature is thus auto-
matically converted to the conductivity
of a freely selectable reference
temperature (usually 20 C or 25 C). The conversion takes place with
the aid of the temperature coefficient
with respect to the reference
temperature .
Conductivity at reference temperature
Temperature coefficient
Conductivity at reference temperature
Conductivity at measuring temperature
Measuring temperature
Reference temperature
Temperature coefficients with respect to the reference temperature
(unit: %/°C).
5.6.3.3.6.5
MET Cond - Stop conditions
Tab: Method
▶ MET Cond ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Entry
25 characters
Conditions for stopping the titration. If several stop conditions are set,
then the criterion which is fulfilled first will stop the titration.
Stop volume
Stop when the entered volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.
Input range
0.00000 - 9,999.99 mL
Default value
100.000 mL
Selection
off