5 Method
tiamo 2.4
■■■■■■■■
895
Device type
Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.
If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.
Selection
756 KF Coulometer | 831 KC Coulometer
Default value
831 KC Coulometer
Dosing device
Dosing device
Selection of the number of the dosing device (dosing or exchange unit)
with which the solution is to be dosed. All the dosing device connections
which are possible with the selected device type are displayed.
Titrando
Selection
1 | 2 | 3 | 4
Default value
1
855
Selection
1 | 2 | 3
Default value
1
Titrino
Selection
internal D0
736, 751, 799
Selection
internal D0 | external D1 | external D2
Default value
internal D0
Solution
Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked. At the start of the
command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.
Entry
24 characters
Selection
'Solution name' | not defined
Default value
not defined