6 Configuration
tiamo 2.4
■■■■■■■■
1429
Calibration range
Minimum and maximum concentration of the calibration solutions. In this
range, the calibration curve depicts a monotonously ascending or
descending function. The value cannot be modified. It is defined from the
method and corresponds to the range between the first and last calibra-
tion points used.
Calibration temperature
Temperature at which the calibration is carried out. In addition, the display
shows whether the temperature was measured with a Pt1000 or an NTC
temperature sensor or entered manually.
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
NOTE
If the Calibration temperature has been entered manually, then the
following parameters will be reset to the default value or updated,
respectively:
■
Calibration temperature = manual
■
Calibration date is updated.
■
Calibration method = manual
■
Determination ID is deleted
■
User is updated
■
Audit Trail entry ensues
Calibration date
Display of date and time of the last calibration entered after each auto-
matic calibration or manual entry.
Calibration method
Display of the method name with which the last calibration has been car-
ried out. If the calibration data has been entered manually, then manual
will appear.
Determination ID
Identification of the determination.
The determination ID can be copied and pasted into another field.
User
Display of the short name of the user logged in during the calibration or
who entered the calibration data manually. If work is not carried out using