8 report template - group field, Group field – Metrohm tiamo 2.4 Manual User Manual
Page 258
4.4 Templates
■■■■■■■■■■■■■■■■■■■■■■
242
■■■■■■■■
tiamo 2.4
Selection
Calculations | Electrode test data | Calibration
data | Curves | Messages | Measuring point
list | Method parameters | Result list | Raw
data (endpoints) | Standard addition | Statis-
tics data (short) | Statistics data (long) | Titra-
tion and measuring parameters | Signature
list Determination | Signature list Method |
Variables | Common variables used | Devices
used | Global variables used | Solutions used |
Sample solution used (TC conductivity) | Sen-
sors used | Used colorimetric sensors | Moni-
toring report
Default value
Calculations
Command name
Entry of the command name for which a measuring point list is to be pro-
duced. With not defined, the lists for all curves present in the determina-
tion will be produced as default.
This parameter appears only when Fixed report = Measuring point
list.
Entry
50 characters
Selection
not defined
Default value
not defined
4.4.1.4.3.8
Report template - Group field
Program window: Database
▶ Report template
A group field is for the purpose of grouping together a variety of fields in
the report template. All of the fields gathered together in a group field
can be moved together as a group. The group field prevents a page break
inside the group field.
The group field always extends across the entire width of a page; only the
upper edge (Y value) and the height of the field can be configured.
NOTE
The following fields, which do not permit page break controls, cannot
be inserted into a group field.
■
Fixed report
■
Curve
■
Calibration curve