Metrohm tiamo 2.4 Manual User Manual
Page 1635
■■■■■■■■■■■■■■■■■■■■■■
8 Manual control
tiamo 2.4
■■■■■■■■
1619
Measuring temperature
Temperature, which can be entered manually. If a temperature sensor is
connected, the temperature is measured continuously and the value
entered will be ignored. The value is used for the temperature correction
of the measurement.
Titrando, 855
Input range
–20.0 - 150.0 °C
Default value
25.0 °C
Titrino
Input range
–170.0 - 500.0 °C
Default value
25.0 °C
I(pol)
The polarization voltage is the voltage applied to the polarizable electrode
during a amperometric measurement.
Titrando, 855
Input range
–125.0 - 125.0 µA (Increment: 0.5)
Default value
5.0 µA
Titrino
Input range
–127 - 127 µA (Increment: 1)
Default value
5 µA
U(pol)
The polarization voltage is the potential applied to the polarizable elec-
trode during an amperometric measurement.
Titrando, 855
Input range
–1250 - 1250 mV (Increment: 25)
Default value
400 mV
Titrino
Input range
–1270 - 1270 mV (Increment: 10)
Default value
400 mV
Reference temperature
The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically con-
verted to this reference temperature (usually 20 °C or 25 °C).
712
Input range
–170.0 - 500.0 °C
Default value
20.0