Metrohm tiamo 2.4 Manual User Manual
Page 1171
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1155
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead
If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.
Selection
Device types | Titrando
Default value
Titrando
Sensor
Measuring input
Selection of the measuring input the sensor is connected to.
Titrando, 867
Selection
1 | 2
Default value
1
855
Selection
1
Default value
1
Titrino
Selection
1 | 2 | diff
Default value
1
Sensor
Selection of a sensor of type pH electrode from the list of sensors avail-
able in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.
Selection
Sensor name | pH electrode
Default value
pH electrode
Temperature measurement
Type of temperature measurement.
Titrando, 855, 867
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.