5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1070
■■■■■■■■
tiamo 2.4
Measurement
Two types of measurement can be selected, Measurement with drift
control or Measurement without drift control.
Selection
Measurement with drift control | Measure-
ment without drift control
Default value
Measurement with drift control
Measurement with drift control
The measurement is carried out drift-controlled if this option is
selected. The measurement is canceled as soon as the defined Signal
drift or the Stop measured value is reached or the defined Max.
waiting time has elapsed. The corresponding parameters are not visi-
ble if this option is disabled.
Measurement without drift control
The measurement is carried out without drift control if this option is
selected. Measurement is continued for as long as one of the two stop
criteria Measuring time or Stop measured value is fulfilled. The
corresponding parameters are not visible if this option is disabled.
Measurement with drift control
If the Measurement with drift control option has been selected, then
the following parameters will be displayed:
Signal drift
The measured value is only accepted if the signal drift defined here has
been fallen short of.
Input range
0.1 - 999,000 mAU/min
Default value
1.0 mAU/min
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift contin-
ues to be checked while the minimum waiting time is elapsing.
Input range
0 - 999,999 s
Default value
0 s
Max. waiting time
If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered, then a waiting time that is suitable for
the drift will be calculated automatically according to the following for-
mula: