5 kfc - titration parameters, Titration parameters – Metrohm tiamo 2.4 Manual User Manual
Page 854
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
838
■■■■■■■■
tiamo 2.4
Delay time
If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time.
Input range
0 - 999 s
Default value
10 s
5.6.3.6.5
KFC - Titration parameters
Tab: Method
▶ KFC ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the running of the titration are defined on this tab.
Extraction time
The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). The entry of an extraction time
may be advisable, e.g. for the titration of sparingly soluble samples.
Input range
0 - 999999 s
Default value
0 s
Temperature
Manually entered titration temperature.
756, 831
Input range
-170.0 - 500.0 °C
Default value
25.0 °C
851, 852
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
756, 831
Input range
1 - 999999 s
Default value
2 s
851, 852
Input range
0.1 - 999999.0 s
Default value
2.0 s