5 Method
tiamo 2.4
■■■■■■■■
1183
Monitoring Temperature
Monitoring Temperature
on | off (Default value: off)
If this option is switched on then the temperature will be monitored and
any exceeded limits will be entered in the measuring point list.
Lower limit
Lower limit of the temperature. When the temperature falls below this
value, the action Lower temperature limit exceeded is triggered.
Titrando, 855
Input range
-20.0 - 150.0 °C
Default value
-20.0 °C
Titrino
Input range
-170.0 - 500.0 °C
Default value
-170.0 °C
Lower hysteresis
Lower hysteresis of the temperature. When the temperature once again
exceeds the lower limit by this hysteresis value, the action Temperature
lower limit ok is triggered.
Titrando, 855
Input range
0.0 - 150.0 °C
Default value
0.2 °C
Upper limit
Upper limit of the temperature. When the temperature exceeds this value,
the action Temperature upper limit exceeded is triggered.
Titrando, 855
Input range
-20.0 - 150.0 °C
Default value
150.0 °C
Titrino
Input range
-170.0 - 500.0 °C
Default value
500.0 °C
Upper hysteresis
Upper hysteresis of the temperature. When the temperature once again
falls below the upper limit by this hysteresis value, the action Tempera-
ture upper limit ok is triggered