5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
988
■■■■■■■■
tiamo 2.4
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
0.5 - 10.0 µA/s
Default value
5.0 µA/s
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is activated, then the associated time and temperature for
the maximum measured value will be interpolated from the measuring
point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
0.5 - 10.0 µA/s
Default value
5.0 µA/s
Break point evaluation
NOTE
This evaluation method is possible only with Titrando, 855 and 867.
Break point evaluation
on | off (Default value: off)
If this option is activated, then sharp, almost 90° changes in directions in
the measurement curve will be evaluated.
EP criterion
Measure of the minimum sharpness of the break point. The smaller the EP
criterion set, the more break points will be found. As this is a relative value
related to the total measured value modification, even small changes in
the measured value can be evaluated as a break point for a small mea-
sured value range.
Input range
0 - 1.0
Default value
0.3