5 Method
tiamo 2.4
■■■■■■■■
837
Selection
maximum
Min. rate
This parameter defines the minimum rate with which iodine is generated
inside the Control range. The smaller the rate the slower the titration
and the higher the accuracy.
756, 831
Input range
0.3 - 999.9 µg/min
Default value
0.3 (slow), 15.0 (optimal), 30.0 (fast) µg/min
Selection
minimum
851, 852
Input range
0.3 - 999.0 µg/min
Default value
0.3 (slow), 15.0 (optimal), 30.0 (fast) µg/min
Selection
minimum
Stop criterion
Stop criterion
Switches off titration when the endpoint has been reached and the stop
criterion has been met.
756, 831
Selection
Drift | rel. drift
Default value
rel. drift
851, 852
Selection
Drift | rel. drift | Time | off
Default value
rel. drift
Stop drift
The titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift.
Input range
1 - 999 µg/min
Default value
5 µg/min
Relative stop drift
The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion = rel. drift.
Input range
0 - 999 µg/min
Default value
5 µg/min