5 Method
tiamo 2.4
■■■■■■■■
1187
Command variables
The following command variables are generated in the method run by the
command DOS U and can be used in formulas under the designation
'Command name.Variable designation':
Identifica-
tion
Description
.BLV
Blank value of the sensor used for the command (only
for ISE sensors)
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has not ever been started
.CONC
Concentration of the solution used for the command
.CYL
Cylinder volume of the exchange or dosing unit used
for the command
.DBL
Total duration for the processing of the command in s
.EME
Final measured value (measured value after processing
of the command) in the unit of the measured value
.ENP
Electrode slope of the sensor used for the command (in
mV for ISE sensors)
.ETE
End temperature (temperature following processing of
the command) in °C
.EVT
End volume (total dosed volume at the end of the com-
mand) in mL
.FIN
Command status; 1 = Command has been ended
at least once, 0, invalid (variable not available) =
Command has never been ended
.IME
Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.CAx
Calculated value × (1...3) for the last measuring point
of the measuring point list
.LP.DVT
dV/dt for the last measuring point in the measuring
point list (SET, KFT, STAT, DOS) or drift for the last
measuring point in in the measuring point list in µg/min
(KFC)