5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1170
■■■■■■■■
tiamo 2.4
Dosing rate in mL/min = dosing volume/(total duration - number of refill-
ings • filling time • 1/60)
5.6.6.3.3
DOS - Track call
Dialog window: Method
▶ DOS pH/DOS U ▶ Monitoring ▶ [New]/[Proper-
ties]
▶ Track call #
Monitoring
Selects the quantity at which a track is to be started when its limit value is
exceeded.
Selection
Measured value | Temperature | Any
Default value
Measured value
Any
Either one of the two sizes.
Limit exceeding
Selects the exceeded limit for which a track is to be started.
Selection
Lower limit | Upper limit | Any | OK
Default value
Any
Any
The track will be started if either the upper or the lower limit is
exceeded.
OK
The track is started when the monitored quantities are once again
within the limit values (including hysteresis).
Track name
Selects the track that is to be started automatically.
Selection
Selection of available tracks
NOTE
If a track is called for that is already running then it will be allowed to
finish and then restarted when it is free again.
5.6.6.3.4
DOS pH
5.6.6.3.4.1
DOS pH - Overview
Dialog window: Method
▶ DOS pH ▶ Properties... ▶ DOS pH - 'Command name'
Command for controlled dosing with measured value pH.
Devices
This command can be executed with the following devices: