5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1126
■■■■■■■■
tiamo 2.4
Designa-
tion
Description
.EME
End measured value (measured value after processing of
the command) in the unit of the measured value
.ETE
End temperature (temperature following processing of
the command) in °C
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command
has never ended
.IME
Initial measured value (measured value before start con-
ditions are processed) in the unit of the measured value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.MEA
Measured value for the last measuring point of the mea-
suring point list in the unit of the measured value
.LP.TEM
Temperature for the last measuring point of the measur-
ing point list in °C
.LP.TIM
Time in s until the last measuring point in the measuring
point list is reached
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off)
.NMP
Number of measuring points in measuring point list
.STY
Type of stop with which the command was stopped: 1
= normal; 0 = manual or after error
5.6.5.9.2
ELT MEAS - General/Hardware
Tab: Method
▶ ELT MEAS ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the device are defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.